Description
NI 6030E/6031E/6032E/6033E
Family Specifications
This document lists the I/O terminal summary and specifications for the devices that make up the NI 6030E family of devices.
This family includes the following devices:
• NI PXI-6030E
• NI PCI-MIO-16XE-10 (NI 6030E)
• NI PCI-6031E
• NI PXI-6031E
• NI PCI-6032E
• NI PCI-6033E
I/O Terminal Summary
Note With NI-DAQmx, National Instruments revised its terminal names so they are easier to understand and more
consistent among NI hardware and software products. The revised terminal names used in this document are usually
similar to the names they replace. For a complete list of Traditional NI-DAQ (Legacy) terminal names and their
NI-DAQmx equivalents, refer to Terminal Name Equivalents of the E Series Help.
Product Dimensions:
NI 6030E/6031E/6032E/6033E
Family Specifications
This document lists the I/O terminal summary and specifications for the devices that make up the NI 6030E family of devices.
This family includes the following devices:
• NI PXI-6030E
• NI PCI-MIO-16XE-10 (NI 6030E)
• NI PCI-6031E
• NI PXI-6031E
• NI PCI-6032E
• NI PCI-6033E
I/O Terminal Summary
Note With NI-DAQmx, National Instruments revised its terminal names so they are easier to understand and more
consistent among NI hardware and software products. The revised terminal names used in this document are usually
similar to the names they replace. For a complete list of Traditional NI-DAQ (Legacy) terminal names and their
NI-DAQmx equivalents, refer to Terminal Name Equivalents of the E Series Help.
Accuracy Information
Transfer Characteristics
Relative accuracy,
or integral nonlinearity (INL) ……… ±0.5 LSB typ,
±1 LSB max
DNL ………………………………………… ±1 LSB max
Monotonicity…………………………….. 16 bits, guaranteed
Offset error
After calibration……………………..305 µV max
Before calibration …………………..20 mV max
Gain error (relative to internal reference)
After calibration……………………..±30.5 ppm max
Before calibration …………………..±2,000 ppm max
Nominal Range (V) Absolute Accuracy Absolute
Accuracy at
Full Scale
(mV)
Positive
Full Scale
Negative
Full Scale
% of Reading Offset
(µV)
Temp Drift
(%/°C) 24 Hours 90 Days 1 Year
10 –10 0.0045 0.0053 0.0062 812.8 0.0001 1.430
10 0 0.0045 0.0053 0.0062 583.9 0.0001 1.201
Note: Accuracies are valid for measurements following an internal E Series calibration. Averaged numbers assume dithering
and averaging of 100 single-channel readings. Measurement accuracies are listed for operational temperatures within ±1 °C
of internal calibration temperature and ±10 °C of external or factory-calibration temperature. NI recommends a one-year
calibration interval. The Absolute Accuracy at Full Scale calculations were performed for a maximum range input voltage
(for example, 10 V for the ±10 V range) after one year, assuming 100 points of averaged data. Go to ni.com/info and enter
info code rdspec for example calculations.
Please make an inquiry for more models:
778039-01 | NI | PXI-6259 | NI |
778271-01 | NI | PXI-6508 | NI |
778927-01 | NI | PXI-6533 | NI |
780040-01 | NI | PXI-8101 | NI |
780587-29 | NI | PXI-8106 | NI |
781093-01 | NI | PXI-8115 | NI |
782857-01 | NI | PXI-8176 | NI |
960903-02 | NI | PXI-8187 | NI |
AT-DIO-32F | NI | PXI-8221 | NI |
AT-GPIB/TNT | NI | PXI8331 | NI |
AT-MIO-16E-2 | NI | PXI-8331/8336 | NI |
BNC-2110 | NI | PXI-8423 | NI |
BNC-2110-777643-01 | NI | PXI-8431-8 | NI |
BNC-2121 | NI | PXI-8432/2 | NI |
CB-50LP | NI | PXI-8461 | NI |
CB-68LP | NI | PXIE-6363 | NI |
CFP-2110 | NI | PXIE-6528 | NI |
CFP-AI-111 | NI | PXIE-8135 | NI |
CFP-AO-200 | NI | PXIE-8431/16 | NI |
CFP-CB-1 | NI | PXI-GPIB | NI |
CFP-CB-3 | NI | PXI-PCI8331 | NI |